Laminar Air Flow

EPS laminar flows are the premium selection for the discerning researcher, offering a combination of value, high quality construction, low operating noise levels, and a wide product range to suit all budgets, from the Research and Industry leader

Horizontal and Vertical Laminar Flow Clean Benches

EPS Laminar Flows are available in both horizontal and vertical flow models. In both models, air is taken in from above the cabinet and passed through a HEPA filter.

Blower System: –   Direct drive, continuous duty ¼ HP with sealed bearings motors selected for energy efficiency, compact design, and flat profile. Completely integrated assembly optimizes motor cooling. All rotating parts balanced for smooth, quiet, vibration-free operation.

Pre filter System: Washable type pleated pre-filter with 90% efficiency is provided.

HEPA Filtration System: – Aluminum framed High efficiency Particulate air (HEPA) filter operate a typical efficiency of 99.97% at 0.3 micron size, provides superior product protection.

 User Interface: – An angled front, curved work surface front edge, and glass sides promote ergonomics. The vertical air flow design minimizes direct airflow which may lead to dry eyes and fatigue on horizontal flow models.

Material of Construction: EPS Models available in laminated finish / 18 gauge Mild steel (MS) material with working area SS non corrosive material.

Control panel: Switch controls for blower, light, outlet and UV are easy to clean and allow airflow to be monitored more accurately.

Product Protection: – Horizontal/Vertical laminar air flow with HEPA filtration, 99.97% at 0.3 micron size, provides superior product protection. 

Air Flow:90 FPM (0.45 m/s) 10 FPM (0.05 m/s) average velocity measured 6 in. (152.4 mm) from the diffuser screen. Uniformity 20% of average or better.20 to 100 fpm. Compared with horizontal flow models, vertical flow clean benches generate less turbulence when large instruments or items are placed in the work zone. 

Key Features:  

  • The blower system maintains airflow as the filter becomes loaded, ensuring optimum efficiency and product protection.
  • EPS products are manufactured for the most demanding laboratory applications.
  • All components are designed for maximum chemical resistance and enhanced durability for a long service life.
  • Each Laminar Air Flow is individually factory tested for safety and performance in accordance with international standards.
laf

Technical Specifications & Ordering Information

Laminar Air Flow-Horizontal

ModelInner working size in( mm)Overall Dimension of Equipment in mm (LXDXH)
Material of Construction
Inner OuterOptional
EPS/LAF-H6002’ x 2’ x 2’660 x 840 x 1740SSMSFully SS
EPS/LAF-H9003’ x 2’ x 2’960 x 840 x 1740SSMSFully SS
EPS/LAF-H12004’ x 2’ x 2’1260 x 840 x 1740SSMSFully SS
EPS/LAF-H18006’ x 2’ x 2’1860 x 840 x 1740SSMSFully SS
Technical Specification
Cleanliness Level CLASS 100
Air Velocity90 FPM±20
Particle Retention 0.3u & above
Accessories Door, UV lamp, Manometer, gas inlet, castors, power-point. HEPA filter, Pre-filter, other standard features.

Technical Specifications & Ordering Information

Laminar Air Flow- Vertical

ModelInner working (size in MM)Overall Dimension of Equipment in mm (LXDXH)
Material of Construction
Inner OuterOptional
EPS/LAF-V6002’ x 2’ x 2’660 x 780 x 1980SSMSFully SS
EPS/LAF-V9003’ x 2’ x 2’960 x 780 x 1980SSMSFully SS
EPS/LAF-V12004’ x 2’ x 2’1260 x 780 x 1980SSMSFully SS
EPS/LAF-V18006’ x 2’ x 2’1860 x 780 x 1980SSMSFully SS
Technical Specification
Cleanliness Level CLASS 100
Air Velocity90 FPM±20
Particle Retention 0.3u & above
Accessories Door, UV lamp, Manometer, gas inlet, castors, power-point. HEPA filter, Pre-filter, other standard features.